TCRT April 2009

category image Volume 8
No. 2 (p 85-176)
April 2009
ISSN 1533-0338
Helical Tomotherapy

Evaluation of Inter-fractional Setup Shifts for Site-specific Helical Tomotherapy Treatments (p. 115-122)

This paper proposes to summarize and analyze the daily patient setup shifts based on megavoltage computed tomography (MVCT) image registration results for Helical TomoTherapy® (HT) treatment. One hundred and fifty-five consecutive treatment plans for a total of 137 patients delivered by the HT unit through one year were collected in this study. The patient data included pelvis (26%), abdomen (23%), lung (21%), head and neck (10%), prostate (8%), and others (12%). All the translational and roll rotational shifts made via auto MVCT and kilovoltage computed tomography (kVCT) image registration were recorded at each fraction. Manual fine-tuning was followed if automatic registration result was not satisfactory. The mean shift ± one standard deviation (1 SD) was calculated for each patient based on the entire treatment course. For each treatment site, the average shift was analyzed as well as displacement in 3D vector. Statistical tests were performed to analyze the relationship of patient-specific, tumor site-specific, and fraction number association with the patient setup shifts. For all the treatment sites, the largest average shift was found in the anterior-posterior direction. The population standard deviations were between 1.2 and 5.6 mm for the X, Y, and Z directions and ranged from 0.2 to 0.6 degrees for the roll rotational correction. The largest standard deviations of the setup reproducibility in X, Y, and Z directions were found in lung patients (4.2 mm), abdomen, lung and spine patients (4.4 mm), and prostate patients (5.6 mm), respectively. The maximum 3D displacement was 10.9 mm for prostate patients? setup. ANOVA tests demonstrated the setup shifts were statistically different between patients even for those that were treated at the same tumor site in the translational directions. No strong correlation between the setup and the fraction number was found. In conclusion, the MVCT guided function in the HT treatment enables us to generate relatively accurate daily setup through registration with KVCT data sets. Our results indicate that lung, prostate, and abdominal patients are more prone to setup uncertainty and should be carefully evaluated.

Key words: Tomotherapy; MVCT; Daily setup; Verification; Positioning.

Lan Lin, Ph.D.1,2,*
Chengyu Shi, Ph.D.1,2
Tony Eng, M.D.2,3
Gregory Swanson, M.D.2,3
Martin Fuss, M.D.4
Niko Papanikolaou, Ph.D.1,2,3

1Department of Radiology
2Cancer Therapy and Research Center
3Department of Radiation Oncology
University of Texas Health Science Center at San Antonio
7703 Floyd Curl Drive
San Antonio, TX 78229, USA
4Department of Radiation Medicine
Oregon Health and Science University
3181 S. W. Sam Jackson Park Rd
Portland, OR 97239, USA

*lanlin@marybird.com

Purchase Downloadable Full-text PDF of Articles

Corporate User

$100.00

University/Academic User

$50.00

Subscription is more cost effective than purchasing PDFs on-the-fly.  Click here for details.

Member Login | Home | Editorial Board | Instructions | Subscribe | Contact Us

Adenine Press, 2066 Central Avenue, Schenectady, NY 12304 USA
phone: 518-456-0784; fax: 518-452-4955; email: info@adeninepress.com
copyright © Adeninepress, All rights reserved.