TCRT April 2008

category image Volume 7
No. 2 (p 91-154)
April 2008
ISSN 1533-0338
Helical Tomotherapy

A Monte-Carlo Derived Dual-Source Model for Helical Tomotherapy Treatment Planning (p. 141-148)

Full Monte Carlo radiation transport simulations of accelerator heads are impractical for routine treatment planning because of the excessive computational burden and memory requirements. To improve the accuracy and efficiency of treatment plans for helical tomotherapy, we have developed a dual-source model to characterize the radiation emitted from the head of a commercial helical tomotherapy accelerator. Percentage depth dose (PDD) and beam profiles computed using the dual-source model with the EGS/BEAMnrc Monte Carlo package agree within 2% of measurements for a wide range of field sizes, which suggests that the proposed dual-source model provides an adequate representation of the tomotherapy head for dose calculations in routine treatment planning.

Key words: Monte Carlo simulation; Photon dosimetry; Source modeling; and Tomotherapy.

Yayun Hsiao, M.S.1
Robert D. Stewart, Ph.D.2,*
X. Allen Li, Ph.D.3,a

1Department of Physics
Purdue University
West Lafayette, IN 47907-2036, USA
2School of Health Sciences
Purdue University
West Lafayette, IN 47907-2051, USA
3Department of Radiation Oncology
Medical College of Wisconsin
Milwaukee, WI 53226, USA
*trebor@purdue.edu

Purchase Downloadable Full-text PDF of Articles

Corporate User

$100.00

University/Academic User

$50.00

Subscription is more cost effective than purchasing PDFs on-the-fly.  Click here for details.

Member Login | Home | Editorial Board | Instructions | Subscribe | Contact Us

Adenine Press, 2066 Central Avenue, Schenectady, NY 12304 USA
phone: 518-456-0784; fax: 518-452-4955; email: info@adeninepress.com
copyright © Adeninepress, All rights reserved.