TCRT August 2006No. 4 (p 299-444) August 2006 ISSN 1533-0338 Resolution Improvement in Positron Emission Tomography Using Anatomical Magnetic Resonance Imaging (p. 311-318)An ideal imaging system should provide information with high-sensitivity, high spatial, and temporal resolution. Unfortunately, it is not possible to satisfy all of these desired features in a single modality. In this paper, we discuss methods to improve the spatial resolution in positron emission imaging (PET) using a priori information from Magnetic Resonance Imaging (MRI). Our approach uses an image restoration algorithm based on the maximization of mutual information (MMI), which has found significant success for optimizing multimodal image registration. The MMI criterion is used to estimate the parameters in the Sharpness-Constrained Wiener filter. The generated filter is then applied to restore PET images of a realistic digital brain phantom. The resulting restored images show improved resolution and better signal-to-noise ratio compared to the interpolated PET images. We conclude that a Sharpness-Constrained Wiener filter having parameters optimized from a MMI criterion may be useful for restoring spatial resolution in PET based on a priori information from correlated MRI.
Keywords: Spatial resolution; Mutual information; Positron emission tomography; and Magnetic resonance imaging. Yong Chu, Ph.D.1,2* 1Tu and Yuen Center for Functional Onco-Imaging Subscription is more cost effective than purchasing PDFs on-the-fly. Click here for details. |
|
Member Login | Home | Editorial Board | Instructions | Subscribe | Contact Us