TCRT April 2007No. 2 (p 57-150) April 2007 ISSN 1533-0338
Stereotactic Radiosurgery Optically Guided Linac Radiosurgery with a Leksell Head Frame as an Adjunct to Gamma Knife™ Treatment (p. 123-126)Because of geometrical limitations in the helmet of the Leksell Gamma Knife™ (Elekta Corp., Atlanta, GA, USA) certain regions within the cranium cannot be targeted for treatment. We describe a method by which lesions in these regions can be treated with the Varian-Zmed stereotactic radiosurgery system utilizing an infrared optical positioning system attached to a Leksell head frame. We have measured the accuracy of the optical tracking system using a phantom attached to a Leksell frame and have determined that the system can target a linear accelerator radiosurgery beam to an accuracy of within 1 millimeter.
Key words: Optical guidance; Gamma Knife; and RadioCameras. Kenneth E. Ekstrand, Ph.D.* Department of Radiation Oncology Subscription is more cost effective than purchasing PDFs on-the-fly. Click here for details. |
|