TCRT February 2007No. 1 (p 1-56) February 2007 ISSN 1533-0338 Breast Skin Doses From Brachytherapy Using MammoSite® HDR, Intensity Modulated Radiation Therapy, and Tangential Fields Techniques (p. 17-22)Skin doses from brachytherapy using MammoSite® HDR, Intensity Modulated Radiation Therapy (IMRT), and conventional tangential fields techniques were compared. For each treatment technique, skin doses were measured using paired thermoluminescent dosimeters placed on the patient?s skin: (i) directly above the balloon catheter during MammoSite® HDR brachytherapy treatments and (ii) 4 cm inside the treatment borders during the IMRT and conventional breast treatments. The mean dose measured was about 58% of the prescription dose for the patients treated using the MammoSite® technique. On the other hand, for patients treated with IMRT and tangential fields, the mean dose was found to be about 69% and 71% of the corresponding prescription dose. This study suggests that in breast cancer radiation treatments the MammoSite® HDR technique reduces skin doses compared to IMRT and tangential field techniques.
PACS numbers: 87.53.Jw, 87.53.Xd, 87.66.Sq Keywords: Breast cancer; Skin dose; MammoSite®; IMRT; and Conventional. Raj N Selvaraj, M.S.1,* 1Department of Radiation Oncology Subscription is more cost effective than purchasing PDFs on-the-fly. Click here for details. |
|
Member Login | Home | Editorial Board | Instructions | Subscribe | Contact Us
Adenine Press, 2066 Central Avenue, Schenectady, NY 12304 USA |
|